범주
연락처
주소:
중국 강소성 무석시 빈호시 건죽로 599호 국가공업설계단지 1-1009, 우편번호: 214062
전화기:Tell: +86 21 51096910
EMAIL:
Email: info@welsobio.com
조사
인큐베이터 셰이커
제품 상세 정보
In the process of cell culture, the growth environment is particularly important, and the culture conditions of various cells are also very different. The incubator shaker provided by Welso can meet the needs of the culture of various cells. Through shaking culture, the cells and bacteria can be reproduced evenly and the culture efficiency is higher. High, and equipped with mobile phone wireless monitoring software for the Internet or local area network, it can wirelessly monitor parameters such as temperature, rotation speed, humidity, light, etc., and can also receive alarm information and cancel the alarm. WIS4000M can be stacked up to three levels, meets the height requirements of laboratory personnel, is easy to operate, saves laboratory space, and has flexible accessory selection to meet various needs of the laboratory.
Features
•10" high resolution LCD touch screen with simultaneous display of all parameters.
•Bi-directional ultraviolet sterilization.
•Slide-out platform for easy cleaning and accessibility of samples.
•Large capacity for batch culture and shaking.
•Optional WiFi and computer monitoring systems allow remote control to ensure safety.
•CO₂ infrared sensor and effective inlet filter.
•Temp. uniformity ≤±0.4°C(@37℃).
Specifications
Model | WIS4000M | WIS4000MC | WIS4005M | WIS4005MC |
Product Name | Stackable Incubator Shaker,Touch Screen | Stackable Incubator Shaker With CO₂ Controller, Touch Screen | Stackable Incubator Shaker,Touch Screen, Shaker Diameter 50mm | Stackable Incubator Shaker With CO₂ Controller, Touch Screen, Shaker Diameter 50mm |
Shaking | ||||
Orbit Diameter | 26mm | 50mm | ||
Speed | 0 (stop), 30 - 300 rpm | |||
Speed Accuracy | ±1rpm | |||
CO₂ | ||||
CO₂ Sensor | NA | IR | NA | IR |
CO₂ Control Range | NA | 0-20% | NA | 0-20% |
CO₂ Control Accuracy | NA | ±0.1% | NA | ±0.1% |
Recover Time After Open Door 30s on Temp | ≤6min(@37℃) | |||
Recover Time After Open Door 30s on CO₂ | NA | ≤6min(@5%) | NA | ≤6min(@5%) |
Temperature | ||||
Temp.Range | Room Temp.Minus 20℃~60℃(lowest Temp. is 4℃) | |||
Temp.Accuracy | 0.1℃ | |||
Temp.Uniformity | ±0.3℃(@37℃) | |||
Heating Power | ≤850w | |||
Cooling Power | 70~230w | |||
Humidity | ||||
Humidity Type | NA | Evapor Naturally | NA | Evapor Naturally |
Max.Humidity | NA | 95% | NA | 95% |
General | ||||
Platform Dimensions | 900*560mm | |||
Capacity | 401L | |||
Max.Load | 20KG | |||
Magnetic Driver | Yes | |||
Programmable Segments | 12 Segments | |||
Illumination | LED Lamps on Left and Right Walls | |||
Disinfection | UV Lamps on Left and Right Walls | |||
Door Open Direction | Downward | |||
Data Display | Temp, Speed | Temp, Speed CO₂, RH | Temp, Speed | Temp, Speed, CO₂, RH |
Display | 10" High-Res Color Touch Screen | |||
Timer | 0-999h59min | |||
USB Export Data | Yes | |||
Data Display on Interface | In 7 Days | |||
Operating Data Storage Cycle | More Than 912 Days | |||
Data Storage Interval | 1min | |||
Data Stored | Time, Temp, Speed, Status | Time, Temp, Speed, Status, CO₂ RH | Time, Temp, Speed, Status | Time, Temp, Speed, Status, CO₂, RH |
Auto-Defrost | Yes | |||
Auto Screen Off | Yes | |||
Auto Screen Lock | Yes | |||
Third Platform Height | 1491mm | |||
Platform with Rails | Yes | |||
Noise Level | ≤55dB(When Compressor Is Running) | |||
Auto-run after Power Recovery | Yes | |||
Power | ≤950w | ≤1000w | ≤950w | ≤1000w |
Power Supply | AC 220V±22V, 50Hz/60Hz | |||
RH Control | NA | Optional(10% Above Ambient Humidity, Up To 90%) | NA | Optional(10% Above Ambient Humidity, Up To 90%) |
Mobile Monitor and Control by WIFI | Optional | |||
PC monitor and Control by RS485 | Optional | |||
Window Cover | Optional | |||
Net Weight | 265kg | 270kg | 265kg | 270kg |
Internal Working Height | 335mm | |||
External Dimensions | 1400*886(600+80 Feet High)mm |
Stack Dimensions
관련 권장 사항